Introduction
Materials and methods
Animals
Data acquisition
Data pre-processing
Registration
Perioperative image guidance
Post-processing
Statistical analysis
Results
Data pre-processing
LV end-diastolic volume (ml) | 128 ± 19 |
LV end-systolic volume (ml) | 75 ± 11 |
LV ejection fraction (%) | 42 ± 1 |
Myocardium volume (ml) | 118 ± 18 |
Infarct volume (ml) | 17 ± 3 |
LV area (cm2) | 101 ± 21 |
Infarct area 0 % transmurality (cm2) | 73.6 ± 19.7 (72.5 ± 5.1 %) |
Infarct area 0–25 % transmurality (cm2) | 7.2 ± 1.0 (7.2 ± 0.5 %) |
Infarct area 25–50 % transmurality (cm2) | 9.5 ± 2.6 (9.6 ± 2.9 %) |
Infarct area 50–75 % transmurality (cm2) | 5.8 ± 2.8 (6.4 ± 4.4 %) |
Infarct area 75–100 % transmurality (cm2) | 4.6 ± 5.2 (4.3 ± 4.3 %) |
Registration
Electromechanical mapping points | |||
---|---|---|---|
Total number of points | 192 | ||
Points per animal | 66.3 ± 14.2 | ||
Distance EMM points to MRI mesh surface (mm) | |||
All EMM points | 3.27 ± 1.93 mm based on 192 points | ||
All EMM points in cine mesh | 3.22 ± 1.86 mm based on 183 points | ||
Points per infarct transmurality area | |||
0 % | 53.3 ± 11.7 | 0.8 ± 0.3 points/cm2
| |
0–25 % | 6.0 ± 3.6 | 0.8 ± 0.5 points/cm2
| |
25–50 % | 6.0 ± 3.0 | 0.7 ± 0.4 points/cm2
| ns (p = 0.96) |
50–75 % | 2.6 ± 1.5 | 0.4 ± 0.1 points/cm2
| |
75–100 % | 4.5 ± 4.1 | 0.9 ± 1.2 points/cm2
|
Post processing
Discussion
Data acquisition and pre-processing
Registration
Post-processing
Clinical implications
Limitations
Conclusion